Blog

A comparative study of shade-matching performance using intraoral scanner, spectrophotometer, and visual assessment | Scientific Reports

Thank you for visiting nature.com. You are using a browser version with limited support for CSS. To obtain the best experience, we recommend you use a more up to date browser (or turn off compatibility mode in Internet Explorer). In the meantime, to ensure continued support, we are displaying the site without styles and JavaScript.

Scientific Reports volume  14, Article number: 23640 (2024 ) Cite this article dental furnace

This study aimed to explore the clinical applicability of the shade-matching function in intraoral scanners. This study measured the tooth colors of maxillary anterior dentitions of 83 adults using visual matching, a spectrophotometer, and a scanner according to two color systems: VITA Classical (VC) and VITA 3D-Master (V3D). Agreement between each method was assessed by weighted Cohen’s kappa coefficient (KW, α = 0.05). For V3D, the overall agreement between the scanner and spectrophotometer (KW = 0.498) was higher than that between the scanner and visual matching (KW = 0.473). Similarly, the agreement between the scanner and spectrophotometer (KW = 0.283) was higher than that between a scanner and visual matching (KW = 0.140) for VC. Regarding tooth position, the highest agreement between the scanner and spectrophotometer was observed on the right central incisor (KW = 0.542) for V3D. Tooth color measurement with a scanner was comparable to that with a spectrophotometer, especially on the central incisors when using the VITA 3D-Master system. A scanner could serve as an alternative to a spectrophotometer for shade selection. However, color matching should still be visually verified.

With recent advancements in scientific technology, the latest versions of intraoral scanners now include built-in functions for measuring tooth color as well as reproducing intraoral structures in three-dimensional (3D) images9,10. The 3Shape TRIOS (3Shape, Copenhagen, Denmark) uses ultrafast optical sectioning, which combines confocal microscopy with structured light projection to obtain in-focus images at different depths, transforming contrast data into 3D surface information pixel by pixel11. A previous study12 reported that artificial intelligence (AI)-based approaches can be used for digital shade matching, leveraging algorithms such as fuzzy logic, backpropagation neural networks, convolutional neural networks, and artificial neural networks to precisely select colors and emulate complex human intelligence. Intraoral scanners may vary in accuracy and precision for tooth shade matching because of variations in light sources, camera sizes, and operating mechanisms13. Moreover, the accuracy and reliability of both spectrophotometers and intraoral scanners can be affected by the operator’s training and experience, with variability in proficiency potentially impacting the results and the reproducibility of the study. Additionally, environmental and lighting conditions can impact the performance of both visual and instrumental color matching. Variations in these conditions during color measurements could lead to inconsistencies in the findings.

Table 1 presents the inter-method agreement based on the VITA 3D-Master shade guide system using the weighted Cohen’s kappa coefficient (KW). The highest agreement was observed between the intraoral scanner (IOS) and the spectrophotometer (SP) (KW = 0.498), followed by the SP and visual color matching (VM) (KW = 0.488). Table 2 shows the inter-method agreement based on the VITA Classical shade guide system using the weighted Cohen’s kappa coefficient (KW). The highest agreement was detected between the IOS and SP (KW = 0.283), followed by the agreement between the IOS and VM (KW = 0.140). Table 3 reveals the match percentages between each method based on the shade guide system. The percentage of color match between a scanner and a spectrophotometer was 46.9% when the tooth color was recorded using the VITA 3D-Master system, compared to 32.3% with the VITA Classical system. Additionally, the percentage of color match between a scanner and visual matching was nearly twice as high with the VITA 3D-Master system compared to the VITA Classical shade guide.

The upper central incisor, lateral incisor, and canine on the right side were denoted as #11, #12, and #13, respectively, while the central incisor, lateral incisor, and canine on the left side were denoted as #21, #22, and #23, respectively. In terms of tooth position, the agreement between the IOS and SP according to the V3D system ranged from the highest to the lowest as follows: #11 (KW = 0.542) > #21 (KW = 0.532) > #12 (KW = 0.439), whereas the agreement between the IOS and VM varied as #21 (KW = 0.446) > #13 (KW = 0.418) > #11 (KW = 0.412). For the VC system, the agreement between IOS and SP ranged from #22 (KW = 0.359) > #21 (KW = 0.333) > #11 (KW = 0.272), and between the IOS and VM, in the order of #11 (KW = 0.209) > #21 (KW = 0.174) > #12 (KW = 0.108) (Table 4). In both shade guide systems, the agreement between the IOS and SP, as well as between the IOS and VM, was higher for the central incisors than for the lateral incisors or canines. The interpretation range of the weighted Cohen’s kappa values was as follows: 0–0.2 (slight), > 0.2 (fair), > 0.4 (moderate), and > 0.6 (substantial)18. When converted to percentages, the range corresponded to 0–0.2 (0–4%), > 0.2 (> 4%), > 0.4 (> 16%), > 0.6 (> 36%).

This study examined the inter-method agreement between an intraoral scanner, a spectrophotometer, and visual matching to estimate tooth colors. The results indicated that the inter-method agreement between the intraoral scanner and spectrophotometer, as well as between the intraoral scanner and visual matching, was moderate when considering both the VITA Classical and VITA 3D-Master shade guide reference systems. However, a higher agreement was observed between the intraoral scanner and spectrophotometer than between the intraoral scanner and visual matching. Because of their ability to detect subtle color differences beyond the discernment of the human eye, a higher agreement may be achieved between both instrumental methods. Therefore, the null hypothesis of this study was rejected, despite the fact that intraoral scanners proved to be reliable and can serve as a reference tool for tooth shade selection.

The clinical applicability of intraoral scanners for measuring tooth shade remains controversial. Floriani et al.15 reported that the accuracy of shade determination using a scanner decreased in aged resin composites, particularly for darker shades. Rutkūnas et al.16 found that the accuracy of shade matching with a scanner was 27.5% based on the VITA Classical shade guide system. They also revealed a visually perceptible color difference between the color values obtained using a scanner and those obtained using a spectrophotometer. In Yoon et al.’s study19, the average color difference values (ΔE) between shades obtained by an intraoral scanner and by a colorimeter exceeded 10, despite the scanner demonstrating high repeatability in color measurement. In contrast, Czigola et al.14 suggested that an intraoral scanner could be used for tooth shade assessment combined with visual matching when referring to the VITA 3D-Master shade guide system, which showed higher repeatability than a spectrophotometer.

Previous studies have also indicated that the choice of shade guide during the scanner setup for shade matching is one of the most commonly reported influencing factors. This study showed greater agreement between the scanner and spectrophotometer when using V3D than when using VC. These results support previous findings13,14,20 suggesting that utilizing VITA 3D-Master settings enhances accuracy, as it is believed to closely align with the capabilities of the human eye. The V3D system offers a more suitable array of natural tooth shades than the VC shade guide, thereby leading to fewer color differences when selecting tooth shades21.

The results of this study revealed that a greater agreement was observed between both electronic methods than between scanning and visual matching. Utilizing a high-definition camera with an LED light, the scanner estimates tooth color using specific software22. Consequently, digital approaches offer enhanced repeatability compared to visual systems. The translucency of the enamel, background color, light intensity, and subjective conditions of the examiner may influence the tooth shade perceived by the human eye, although visual matching is still considered the standard technique owing to its simplicity and low cost23. In this study, the maxillary anterior teeth were chosen for color measurement because of their prominent visibility during smiling and significant aesthetic influence. In terms of tooth position, a greater agreement was observed between each pair of methods for the central incisors. This could be attributed to their wide and even buccal surfaces, which reduce the edge-loss effect compared to other teeth24.

Spectrophotometers have demonstrated both repeatability and accuracy in color selection across multiple studies2,15,24. However, these devices are typically not readily available in routine dental practice because of their high costs. In this study, the inter-method agreement between the intraoral scanner and spectrophotometer was moderate. Thus, results regarding whether an intraoral scanner with an integrated shade-taking function can effectively substitute for spectrophotometers are inconclusive. Yilmaz et al.25 suggested that objective instrumental measurement methods could be complemented by subjective visual methods, potentially yielding superior results when used together. However, employing a scanner for shade determination offers advantages in terms of ease of use, time and cost savings, scanning of a wider area with greater susceptibility, and lower operating sensitivity. According to the findings of this study, a scanner could serve as an alternative to a spectrophotometer for shade selection; however, color matching should still be visually verified. Hence, the results of this study affirm the effectiveness of using an intraoral scanner to determine tooth color, although further refinement of computer-based systems is warranted.

The limitations of this study include the lack of standardization of the operating variables of the scanning system, such as scan angle, time, distance, rescanning time, light source, and the presence of shadows from the surrounding tissue. Further research should aim to assess the effects of each variable on color matching using intraoral color scanners. In addition, because the scanner does not provide L*, a*, and b* color values, comparing the color scales obtained using each method is challenging. However, as the technology continues to evolve, improvements in the accuracy and precision of intraoral scanners for tooth color measurement could significantly impact study outcomes. Another limitation is that it included only 83 adult patients from a single dental hospital, which may affect the generalizability of the findings. A more diverse and larger sample size would enhance the validity and applicability of the results to a broader population. Additionally, the visual color matching was performed by a single prosthodontist with over 20 years of experience. The results may be influenced by the practitioner’s unique skills and experience, potentially introducing bias. Including multiple practitioners with varying levels of experience could provide a more comprehensive evaluation of the visual matching method, reducing the impact of subjectivity inherent in visual matching and color perception.

Moreover, the subjective color measurements were taken under natural light with participants seated in the same dental chair, but variations in natural light intensity throughout the day or minor differences in participant positioning could affect the results. Controlled lighting conditions might provide more consistent outcomes. Additionally, while the white balance calibration of the spectrophotometer and the calibration of the intraoral scanner were performed before each measurement, slight variations in calibration processes or operator techniques could still impact the consistency and reliability of the measurements. Lastly, tooth color was obtained only from the middle third of the labial surfaces of specific teeth in this study. Since different parts of the same tooth can exhibit varying shades, assessing only one area may not accurately represent the overall tooth color.

The intraoral scanner in this study effectively matched the spectrophotometer for shade selection using the VITA 3D-Master system, but visual confirmation remains the most reliable method for the best clinical results.

This study was designed to evaluate the agreement between digital shade matching and traditional methods (visual matching and spectrophotometric analysis) using two commonly employed shade guide systems: VITA Classical (VC) and VITA 3D-Master (V3D). The research adhered to the ethical principles outlined in the Declaration of Helsinki and received approval from the Ethics Committee of Ajou University Institutional Review Board (No. AJOUIRB-OB-2023-114). Written informed consent was obtained from all participants prior to their inclusion in the study.

A total of 83 adult patients (44 men and 39 women) from Ajou University Dental Hospital participated in this study. Inclusion criteria required that participants had healthy maxillary central incisors, lateral incisors, and canines without structural defects, discoloration, or restorations to ensure standardized conditions for color matching. However, this exclusion criterion may limit the applicability of the findings to teeth with similar conditions, which are frequently encountered in clinical practice.

Color measurements were obtained from the middle-third of the labial surfaces of 498 teeth (maxillary central incisors, lateral incisors, and canines). To minimize environmental variables, all participants were seated in the same dental chair under natural light, with their heads stabilized on headrests. Foreign substances were removed from the tooth surfaces using gauze before assessments.

The subjective visual color matching (VM) was performed by a prosthodontist with over 20 years of clinical experience using the VC and V3D shade guide systems (Vita Zahnfabrik, Bad Säckingen, Germany) (Fig. 1). The prosthodontist selected shades for each tooth within 10 s to reduce errors related to contrast, residual effects, and eye fatigue26. However, despite these precautions, the inherent subjectivity of visual matching presents a risk of individual bias.

(A) VITA Classical shade guide system; (B) VITA 3D-Master shade guide system; and (C) visual color matching with a shade tab.

Objective color measurements were conducted using the VITA Easyshade V intraoral spectrophotometer (SP) (Vita Zahnfabrik, Bad Säckingen, Germany) (Fig. 2). The spectrophotometer was calibrated with white balance before each use to ensure accuracy. The operator, who was trained in the use of the spectrophotometer, placed the measurement tip perpendicular to the middle segment of the target tooth under consistent conditions27. The colors were recorded as VC and V3D values (Fig. 2C).

(A) VITA Easyshade V spectrophotometer; (B) Shade determination of natural teeth with VITA Easyshade V; and (C) Output of the tooth shade in the VITA Classical and VITA 3D-Master systems.

Digital shade matching was performed using the Medit i700 wireless intraoral scanner (IOS) (Medit, Seoul, Korea) (Fig. 3A). The Medit i700 wireless, released in 2022 by the Korean digital imaging company Medit, is the latest model in their line of intraoral scanners. As the successor to the Medit i500, it features updated software for enhanced performance including tooth shade matching capabilities. The scanner was calibrated prior to each session. Real-time scan data were acquired from the test teeth, and the color measurement mode of the IOS system was used to match the VC and V3D shade systems. The optimal shade, marked with a star among the three recommended colors, was selected for analysis (Fig. 3B and C).

(A) The intraoral scanner tested in this study (Medit i700, Seoul, Korea); Screenshot of the digital color matching on the left central incisor according to (B) VITA Classical, and (C) VITA 3D-Master shade guide systems.

The null hypothesis of this study was that there would be no significant difference in the shade matching results between the digital (IOS), visual (VM), and spectrophotometric (SP) methods when using the VC and V3D shade guide systems. To test this hypothesis, inter-method agreement was evaluated using weighted Cohen’s kappa statistic, with a significance level set at α = 0.05.

All statistical analyses were conducted using R software (version 4.2.3, https://www.r-project.org/). The inter-method agreement for VC and V3D shade systems was determined using weighted Cohen’s kappa to quantify the level of agreement between the digital scanner, visual matching, and spectrophotometric methods. This approach allowed for an assessment of the consistency and reliability of each method in determining tooth color.

However, despite these precautions, the subjective nature of visual matching inherently carries the risk of individual bias.

The datasets generated during the current study are available from the corresponding author (H.-K.K.) on reasonable request.

Tabatabaian, F., Beyabanaki, E., Alirezaei, P. & Epakchi, S. Visual and digital tooth shade selection methods, related effective factors and conditions, and their accuracy and precision: A literature review. J. Esthet. Restor. Dent. 33, 1084–1104. https://doi.org/10.1111/jerd.12816 (2021).

Morsy, N. & Holiel, A. A. Color difference for shade determination with visual and instrumental methods: A systematic review and meta-analysis. Syst. Rev. 12, 95. https://doi.org/10.1186/s13643-023-02263-9 (2023).

Article  PubMed  PubMed Central  Google Scholar 

Alnusayri, M. O., Sghaireen, M. G., Mathew, M., Alzarea, B. & Bandela, V. Shade selection in esthetic dentistry: A review. Cureus. 14, e23331. https://doi.org/10.7759/cureus.23331 (2022).

Article  PubMed  PubMed Central  Google Scholar 

Jouhar, R. et al. Analysis of shade-matching ability in dental students: A comparative study under clinical and correcting light conditions. BMC Med. Educ. 24. https://doi.org/10.1186/s12909-024-05146-2 (2024).

Parameswaran, V., Anilkumar, S., Lylajam, S., Rajesh, C. & Narayan, V. Comparison of accuracies of an intraoral spectrophotometer and conventional visual method for shade matching using two shade guide systems. J. Indian Prosthodont. Soc. 16, 352–358. https://doi.org/10.4103/0972-4052.176537 (2016).

Article  PubMed  PubMed Central  Google Scholar 

Ebeid, K., Sabet, A., El Sergany, O. & Della Bona, A. Accuracy and repeatability of different intraoral instruments on shade determination compared to visual shade selection. J. Esthet. Restor. Dent.Bold">34, 988–993. https://doi.org/10.1111/jerd.12884 (2022).

Kim, H. K. Evaluation of the repeatability and matching accuracy between two identical intraoral spectrophotometers: An in vivo and in vitro study. J. Adv. Prosthodont. 10, 252–258. https://doi.org/10.4047/jap.2018.10.3.252 (2018).

Article  PubMed  PubMed Central  Google Scholar 

Kim, H. K. A study on the color distribution of natural teeth by age and gender in the Korean population with an intraoral spectrophotometer. J. Esthet. Restor. Dent. 30, 408–414. https://doi.org/10.1111/jerd.12424 (2018).

Shah, P. et al. Investigating working practices of dentists on shade taking: Evidence based good practice versus observed practice. J. Dent. 97, 103341. https://doi.org/10.1016/j.jdent.2020.103341 (2020).

Article  CAS  PubMed  Google Scholar 

Mehl, A., Bosch, G., Fischer, C. & Ender, A. In vivo tooth-color measurement with a new 3D intraoral scanning system in comparison to conventional digital and visual color determination methods. Int. J. Comput. Dent. 20, 343–361 (2017).

Amornvit, P., Rokaya, D., Peampring, C. & Sanohkan, S. Confocal 3D optical intraoral scanners and comparison of image capturing accuracy. Comput. Mater. Contin. 66https://doi.org/10.32604/cmc.2020.011943 (2021).

Shetty, S., Gali, S., Augustine, D. & Sv, S. Artificial intelligence systems in dental shade-matching: A systematic review. J. Prosthodont.https://doi.org/10.1111/jopr.13805 (2023).

Akl, M. A., Mansour, D. E. & Zheng, F. The role of Intraoral Scanners in the shade matching process: A systematic review. J. Prosthodont. 32, 196–203. https://doi.org/10.1111/jopr.13576 (2023).

Czigola, A. et al. Comparing the effectiveness of shade measurement by intraoral scanner, digital spectrophotometer, and visual shade assessment. J. Esthet. Restor. Dent. 33, 1166–1174. https://doi.org/10.1111/jerd.12810 (2021).

Floriani, F. et al. A comparative study of shade-matching reproducibility using an intraoral scanner and a spectrophotometer. Dent. J. (Basel). 12, 62. https://doi.org/10.3390/dj12030062 (2024).

Rutkūnas, V., Dirsė, J. & Bilius, V. Accuracy of an intraoral digital scanner in tooth color determination. J. Prosthet. Dent. 123, 322–329. https://doi.org/10.1016/j.prosdent.2018.12.020 (2020).

Huang, M. et al. Evaluation of accuracy and characteristics of tooth-color matching by intraoral scanners based on Munsell color system: An in vivo study. Odontology. 110, 759–768. https://doi.org/10.1007/s10266-022-00694-9 (2022).

McHugh, M. L. Interrater reliability: The kappa statistic. Biochem. Med. (Zagreb). 22, 276–282 (2012).

Yoon, H. I. et al. A study on possibility of clinical application for color measurements of shade guides using an intraoral digital scanner. J. Prosthodont. 27, 670–675. https://doi.org/10.1111/jopr.12559 (2018).

Revilla-León, M., Methani, M. M. & Özcan, M. Impact of the ambient light illuminance conditions on the shade matching capabilities of an intraoral scanner. J. Esthet. Restor. Dent. 33, 906–912. https://doi.org/10.1111/jerd.12662 (2021).

Rao, D. & Joshi, S. Evaluation of natural tooth color space of the Indian population and its comparison to manufacturer’s shade systems. Contemp. Clin. Dent. 9, 395–399. https://doi.org/10.4103/ccd.ccd_144_18 (2018).

Article  PubMed  PubMed Central  Google Scholar 

Michou, S. et al. Automated caries detection in vivo using a 3D intraoral scanner. Sci. Rep. 11, 21276. https://doi.org/10.1038/s41598-021-00259-w (2021).

Article  ADS  CAS  PubMed  PubMed Central  Google Scholar 

Sirintawat, N., Leelaratrungruang, T., Poovarodom, P., Kiattavorncharoen, S. & Amornsettachai, P. The accuracy and reliability of tooth shade selection using different instrumental techniques: An in vitro study. Sensors (Basel). 21, 7490. https://doi.org/10.3390/s21227490 (2021).

Article  ADS  PubMed  Google Scholar 

Hampé-Kautz, V., Roman, T., Schwob, T., Cournault, B. & Etienne, O. In-vivo repeatability of three intra-oral spectrophotometers. J. Esthet. Restor. Dent. 36, 520–526. https://doi.org/10.1111/jerd.13182 (2024).

Yilmaz, B., Irmak, O. & Yaman, B. C. Outcomes of visual tooth shade selection performed by operators with different experience. J. Esthet. Restor. Dent. 31, 500–507. https://doi.org/10.1111/jerd.12507 (2019).

Özat, P. B., Tuncel, İ. & Eroğlu, E. Repeatability and reliability of human eye in visual shade selection. J. Oral Rehabil. 40, 958–964. https://doi.org/10.1111/joor.12103 (2013).

Dozic, A., Kleverlaan, C. J., Aartman, I. H. & Feilzer, A. J. Relation in color of three regions of vital human incisors. Dent. Mater. 20, 832–838. https://doi.org/10.1016/j.dental.2003.10.013 (2004).

This study was funded by the National Research Foundation of Korea (NRF) grant funded by the Korean Government (MSIT: Ministry of Science and ICT; Grant Number: NRF-2022R1F1A1067929). The authors have no financial interests in the comparisons of the materials included in this article.

Department of Perio-Prosthodontic Implantology, Graduate School of Clinical Dentistry, Ajou University, Suwon, Republic of Korea

Joo-Hyun Lee & Hee-Kyung Kim

Department of Prosthodontics, Institute of Oral Health Science, Ajou University School of Medicine, Suwon, Republic of Korea

You can also search for this author in PubMed  Google Scholar

You can also search for this author in PubMed  Google Scholar

Conceptualization, H.-K.K.; methodology, H.-K.K.; investigation, H.-K.K., and J.-H.L.; formal analysis, H.-K.K.; data curation, H.-K.K., and J.-H.L.; software, H.-K.K., and J.-H.L.; writing–original draft preparation, H.-K.K., and J.-H.L.; writing–review and editing, H.-K.K.; funding acquisition, H.-K.K. All authors have read and agreed to the published version of the manuscript.

The authors declare no competing interests.

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Lee, JH., Kim, HK. A comparative study of shade-matching performance using intraoral scanner, spectrophotometer, and visual assessment. Sci Rep 14, 23640 (2024). https://doi.org/10.1038/s41598-024-74354-z

DOI: https://doi.org/10.1038/s41598-024-74354-z

Anyone you share the following link with will be able to read this content:

Sorry, a shareable link is not currently available for this article.

Provided by the Springer Nature SharedIt content-sharing initiative

Scientific Reports (Sci Rep) ISSN 2045-2322 (online)

5 axis dental milling machine 1 Sign up for the Nature Briefing newsletter — what matters in science, free to your inbox daily.